Lubrication & Maintenance of a Handpiece

Handpiece Maintenance

Steps to Maintain Your Handpieces

​​

  1. Sterilize each Handpiece after every use as required.  Do not exceed 135c degrees. After cycle is complete allow Handpiece to cool down to room temperature.

  2. Clean exterior of Handpiece by using alcohol or distilled water with a fine wire brush to remove contaminates.

  3. Lubricate Handpiece by placing 2-3 drops of oil into the air intake hole as per diagram above.

  4. Install bur into handpiece and let run for 10-15 seconds, this will remove any excessive oil and flush out any debris within the Handpiece.

**WARNING**  Do Not Operate Handpiece Without a Bur Installed.